Study on the simultaneous determination of volatile phenol and cyanide in the tap water by flow injection analysis
CSTR:
Author:
Affiliation:

(Lucheng District Center for Disease Control and Prevention,Zhejiang Wenzhou 325000,China)

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective To establish a flow injection analysis method for simultaneous determination of volatile phenol and cyanide in tap water. Methods The tap water was pretreated by a mixed solution of EDTA and ascorbic acid (20.0 g/L), and the flow injection analyzer was used for simultaneous determination of volatile phenol and cyanide. Results The result show that the mixed solution of EDTA and ascorbic acid (20.0 g/L) of 0.5 ml in 50 ml tap water was the best pretreatment. The recovery rate of the tap water was 90.0%-105.0%, and the relative standard deviation was 0.39%-5.85%. The detection limits of volatile phenol and cyanide were 0.000 2 and 0.000 4 mg/L. Conclusion This method showed good precision, high accuracy, and met the requirements of quality control.

    Reference
    Related
    Cited by
Get Citation

ZHANG Lian-qun, ZHANG Wen-zhu, HE Chun-ding. Study on the simultaneous determination of volatile phenol and cyanide in the tap water by flow injection analysis[J].中国食品卫生杂志,2018,30(1):49-53.