Determination of gross alpha radioactivity and gross beta radioactivity in food at the same time through comparative measurement method
Author:
Affiliation:

(Nantong Center for Disease Control and Prevention,Jiangsu Nantong 226007,China)

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    Objective A method was developed for measuring gross alpha radioactivity and gross beta radioactivity at the same time in food. Methods Sample was prepared after drying, coking, ashing, and the counting rate of gross alpha and beta in sample was measured through the low background alpha and beta measurement system, and quantitative comparison was made with the standard source measurement efficiency. Results The detection limit of gross alpha was 0.160-5.31 Bq/kg (fresh weight). The recovery was 82.7%-101.8%. RSD was 5.65%-7.26% (n=5). The detection limit of gross beta was 0.475-9.29 Bq/kg (fresh weight). The recovery was 88.8%-114.8%. RSD was 2.23%-9.19% (n=5). Conclusion This method could determine the gross alpha radioactivity and gross beta radioactivity in food at the same time, simply, efficiently, and is suitable for total radiation monitoring of the batch food sample.

    Reference
    Related
    Cited by
Get Citation

GU Jun, HE Zhi-min, YE Qing-hua, ZHANG Wei-bing. Determination of gross alpha radioactivity and gross beta radioactivity in food at the same time through comparative measurement method[J].中国食品卫生杂志,2018,30(1):46-49.

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:July 05,2017
  • Revised:
  • Adopted:
  • Online: April 02,2018
  • Published: