Application and discussion of the vacuum acid-driven system for heavy metal detection in foods
CSTR:
Author:
Affiliation:

(Ezhou Center for Disease Control and Prevention,Hubei Ezhou 436099, China)

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    To improve the efficiency of digestion of the sample and explore the possibility to replace the traditional acid-driven with efficient vacuum acid-driven method for heavy metal detection.Methods 3 samples of 7 typical categories were detected by graphite furnace atomic absorption method for lead, cadmium, copper, chromium and nickel, traditional acid-driven and vacuum acid-driven method were applied. Results There was no significant difference between two acid-driven methods, and the RSD of vacuum acid-driven method was 0.04%-0.95%. Conclusion Compared with traditional method, the vacuum acid-driven method could effectively reduce the digestion time as well as achieve better precision and accuracy, and it could be used for heavy metals detection.

    Reference
    Related
    Cited by
Get Citation

SUN Qiang, ZENG Jing, YANG Ai-lian, YU Shuang, HUANG Wei-dong, YANG Yun-fen, WU Li-jun. Application and discussion of the vacuum acid-driven system for heavy metal detection in foods[J].中国食品卫生杂志,2016,28(1):61-64.

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:June 02,2015
  • Revised:
  • Adopted:
  • Online: February 19,2016
  • Published:
Article QR Code