Determination of vitamin D3 in calcium tablet with microwave extraction by RP-HPLC
CSTR:
Author:
Affiliation:

(Liaoning Institute for Food Control,Liaoning Shenyang 110000,China)

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    To develop a RP-HPLC method for determination of vitamin D3 in calcium tablet with microwave extraction. Methods Vitamin D3 was extracted by microwave in the ethanol. The HPLC analysis was carried out on a C18 column (4.6 mm×150 mm,5 μm) with ethanol at the flow rate of 1 ml/min. The detection wavelength was set at 264 nm. Results The calibration curves were linear within the range of 0-4 μg/L (r=0.999 8) , and the detection limit was 0.02 μg/L.Conclusion The recoveries for vitamin D3 were between 98.8%-100.0%, and RSD was 0.4%. This method was simple, accurate and practical for the detection of vitamin D3 in health food.

    Reference
    Related
    Cited by
Get Citation

ZHAO Fei, GAO Guang-hui, WANG Feng-jiao. Determination of vitamin D3 in calcium tablet with microwave extraction by RP-HPLC[J].中国食品卫生杂志,2015,27(1):30-33.