Rapid screening of cadmium in rice by X-ray fluorescence spectrometer
DOI:
Author:
Affiliation:

(Food Safety Production Engineering Research Center of Hunan Province,Hunan Changsha 410013,China)

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    A method of X-ray fluorescence spectrometer was established for rapid screening of cadmium in rice.MethodsThe rice was detected by X-ray fluorescence spectrometer after crushed.ResultsThe limit of detection and quantification of detection were 0.035 and 0.094 mg/kg, the standard deviation of accuracy was 0.015 5 and the value of t-test was 0.30. Chi-square value of repeatability was 5.32, and the chi-square value of stability was 11.07 within 12 h. The standard deviation was 0.021 3 between two instruments.ConclusionThe results indicated that the X-ray fluorescence spectrometer could meet the standard requirements of cadmium in rice. The accuracy, repeatability and stability of instrument could meet the requirement of rapid screening. A new rapid screening method for cadmium in ricewas developed. It provided technical support for purchasing grain in grassroots grain depot, acceptance check and supervision of food raw material.

    Reference
    Related
    Cited by
Get Citation

PENG Xin-kai, WANG Hui, YUAN Liang-jing. Rapid screening of cadmium in rice by X-ray fluorescence spectrometer[J].中国食品卫生杂志,2014,26(6):551-554.