Determination of bisphenol A in styrofoam fast food boxes by gas chromatography-mass spectrometry with microwave digestion and solid phase extraction
DOI:
CSTR:
Author:
Affiliation:

(Wuhu center for Disease Control and Prevention,Anhui Wuhu 241000,China)

Clc Number:

Fund Project:

  • Article
  • |
  • Figures
  • |
  • Metrics
  • |
  • Reference
  • |
  • Related
  • |
  • Cited by
  • |
  • Materials
  • |
  • Comments
    Abstract:

    To establish a method of gas chromatography-mass spectrometry with microwave digestion and solid phase extraction to determine bisphenol A in styrofoam fast food boxes. Methods 0.2 g styrofoam fast food boxes was digested in microwave tank, then cleaned by the cartridge (C18, 500 mg,6 ml) and derivated by 100 μl BSTFA-TMCS (99∶1, V/V). Finally, 1 μl solution was injected into GC-MS instrument. Results The linearity range was 0.305 to 3.05 mg/L, r2=0.991. The average recoveries ranged from 90% to 101%, and the coefficient of variation varied between 2.6% to 5.0%. Detection limit for bisphenol A was 15 μg/kg in 0.2 gram sample. Conclusion The method could be used in determination of bisphenol A in styrofoam fast food boxes for its good precision of qualification and quantification.

    Reference
    Related
    Cited by
Get Citation

LI Yong-jing. Determination of bisphenol A in styrofoam fast food boxes by gas chromatography-mass spectrometry with microwave digestion and solid phase extraction[J].中国食品卫生杂志,2013,25(4):344-347.

Copy
Share
Article Metrics
  • Abstract:
  • PDF:
  • HTML:
  • Cited by:
History
  • Received:January 17,2013
  • Revised:
  • Adopted:
  • Online: September 08,2013
  • Published:
Article QR Code