Objective To investigate and optimize the experiment conditions of determining lead and cadmium with antimony film-plated electrode.Methods Antimony film-plated gold electrode was prepared by in-situ plating for the determination of trace lead and cadmium by differential pulse stripping voltammetry.Results Lead and cadmium displayed well defined peaks in the antimony film-plated electrode by differential pulse stripping voltammetry;and the stripping potential were-0.58 V and-0.81 V for lead and ...
Reference
Related
Cited by
Get Citation
Gong Weilei, Yang Jinling, Wang Changqin. In-situ plating antimony film on gold electrode for the determination of lead and cadmium by differential pulse stripping voltammetry[J].中国食品卫生杂志,2011,23(6):524-527.